The U.S. Industrial Control Systems Cyber Emergency Response Team (ICS-CERT) and Philips have warned healthcare providers that one of the company’s radiation dose management tools is affected by potentially serious vulnerabilities.
Launched in November 2014, the Philips DoseWise Portal (DWP) enables healthcare providers to record, analyze and monitor imaging radiation doses for patients and clinicians across multiple diagnostic settings.
Versions 1.1.7.333 and 2.1.1.3069 of DWP are affected by a couple of vulnerabilities that allow a remote attacker to gain access to the application database and the patient health information it stores.
One of the flaws, tracked as CVE-2017-9656 and classified as “critical severity,” exists due to the use of hardcoded credentials for a database account. The second vulnerability, CVE-2017-9654, is an issue related to login credentials being stored in clear text in backend system files.
“For an attacker to use or exploit these vulnerabilities to access the underlying DWP database, elevated privileges are first required in order for an attacker to access the web application backend system files that contain the hard-coded credentials,” Philips said in its advisory.
“Successful exploitation may allow a remote attacker to gain access to the database of the DoseWise Portal application which contains patient health information (PHI). Potential impact could include compromise of patient confidentiality, system integrity, and/or system availability,” the company added.
Related: Learn More at SecurityWeek’s 2017 ICS Cyber Security Conference
Philips said it was not aware of any attacks exploiting these vulnerabilities. The company pointed out that the product is classified as a low-safety-risk medical device.
The vendor expects to release an update and new documentation to address these issues later this month. Patches will be included in version 2.1.2.3118 and for users of version 1.1.7.333 Philips will reconfigure the DWP installation to change and encrypt passwords.
In the meantime, the company has advised users to ensure that security best practices are implemented across their network, and that port 1433 is blocked, except for cases where a separate SQL server is used.
Philips also issued security alerts recently to warn its customers about the NotPetya and WannaCry attacks. The company informed organizations that some products had been affected by the Windows vulnerability exploited by these pieces of malware.
Several other medical device manufacturers also issued warnings regarding these malware attacks.
Related: Hundreds of Flaws Found in Philips Healthcare Product
Related: Researchers Uncover Privilege Escalation Bug in Philips Medical Devices

Eduard Kovacs (@EduardKovacs) is a contributing editor at SecurityWeek. He worked as a high school IT teacher for two years before starting a career in journalism as Softpedia’s security news reporter. Eduard holds a bachelor’s degree in industrial informatics and a master’s degree in computer techniques applied in electrical engineering.
More from Eduard Kovacs
- Vulnerabilities in Honda eCommerce Platform Exposed Customer, Dealer Data
- Barracuda Urges Customers to Replace Hacked Email Security Appliances
- Google Patches Third Chrome Zero-Day of 2023
- ChatGPT Hallucinations Can Be Exploited to Distribute Malicious Code Packages
- AntChain, Intel Create New Privacy-Preserving Computing Platform for AI Training
- Several Major Organizations Confirm Being Impacted by MOVEit Attack
- Verizon 2023 DBIR: Human Error Involved in Many Breaches, Ransomware Cost Surges
- Ransomware Group Used MOVEit Exploit to Steal Data From Dozens of Organizations
Latest News
- Consolidate Vendors and Products for Better Security
- Pharmaceutical Giant Eisai Takes Systems Offline Following Ransomware Attack
- Vulnerabilities in Honda eCommerce Platform Exposed Customer, Dealer Data
- North Korean Hackers Blamed for $35 Million Atomic Wallet Crypto Theft
- Cisco Patches Critical Vulnerability in Enterprise Collaboration Solutions
- Barracuda Urges Customers to Replace Hacked Email Security Appliances
- Android’s June 2023 Security Update Patches Exploited Arm GPU Vulnerability
- BBC, British Airways, Novia Scotia Among First Big-Name Victims in Global Supply-Chain Hack